and I have read your paper[Metabolite Spectral Accuracy on Orbitraps] ,but I still may not solve two isotopes meantimes.
Isotope: [13C5, 15N2]-glutamine
Should I change the correction matrix and purity matrix both,and if I can know the details of labling like the picture following,
do I need to extend the correction matrix? Or just integrate the data to the form like(M+0, M+1,M+2......M+n)?
I sincerely look forward to your reply,Thanks a lot!
The text was updated successfully, but these errors were encountered:
AccuCor does not support the data with more than one labeled isotope.
You need to construct a joint matrix of C and N as the tracer elements and a joint matrix for the isotope purity correction.
Depending on the molecular mass of the metabolite and the resolution, the joint matrix might be ill-conditioned. This is why the correction of dual tracer data is tricky.
for example

this is my raw data
and I have read your paper[Metabolite Spectral Accuracy on Orbitraps] ,but I still may not solve two isotopes meantimes.

Isotope: [13C5, 15N2]-glutamine
Should I change the correction matrix and purity matrix both,and if I can know the details of labling like the picture following,
do I need to extend the correction matrix? Or just integrate the data to the form like(M+0, M+1,M+2......M+n)?
I sincerely look forward to your reply,Thanks a lot!
The text was updated successfully, but these errors were encountered: