Last updated by Ryan Snow on 2/1/23. The following protcol was designed and adapted from the Agilent D1000 Protocol. Please refer to the manufacturer's kit guide for additional protocol information.
This protocol is intended for Genomic DNA.
-
Remove samples and reagents from storage
- Allow sample to acclimate to RT for 30 minutes
-
Record sample and reagent record log form(s) attached
- Ladder (Φ) will always be in position A1
- Store in lab notebook
-
Launch TapeStation Controller software
-
Record sample and reagent information in TapeStation Controller software
- Ladder (Φ) will always be in position A1
-
Flick ScreenTape(s) to remove any bubbles if necessary
-
Load TapeStation deck with consumables
- ScreenTape(s)
- First ScreenTape will be placed into the ScreenTape nest
- Additional ScreenTape(s) will be placed in the ScreenTape rack leaving room for the first ScreenTape
- Tips
- ScreenTape(s)
-
Vortex and spin samples
-
Label tube strip(s) and/or 96-well plates
- Ensure good documentation practices to track sample position
-
Prepare TapeStation Ladder in position A1 (3 uL Genomic Sample Buffer:1 uL Genomic Ladder)
- For 1 ScreenTapes; load 3 uL of Genomic Sample Buffer and 1 uL Genomic Ladder
- For 2 ScreenTapes; load 6 uL of Genomic Sample Buffer and 2 uL Genomic Ladder
- For 3+ ScreenTapes; load 15 uL of Genomic Sample Buffer and 5 uL Genomic Ladder
-
Per sample, load 3 uL Genomic Sample Buffer and 1 uL of DNA sample in planned position(s)
- Apply lid to sample tube strip and/or plate after adding sample
-
Vortex at 2000 rpm for 1 minute using the IKAMS3 vortexer
- Located near the TapeStation
-
Spin down for 1 minute to remove sample and ladder from lid
-
Load TapeStation instrument
- Ladder must be in position A1
-
Remove sample lid before closing and running instrument
- Verify sample presence
- Close TapeStation lid
-
On the TapeStation computer click Start
- Confirm information in pop-up dialogue box
- Move on to Data Analysis after run is complete
-
Follow instructions in Run Finished dialogue box
- Label and store any partially used ScreenTapes as such
- Discard used tips
- Best practice to replace tips after use
- Remove samples from TapeStation
- Store ScreenTape(s) in Used ScreenTapes container
-
Save and export data using the following naming convention: [Initials/Personal Identifier][Relevant Run Info]-[MMDDYY]
Ex. for Ryan Snow Test Run on Jan 18th, 2023 SnTR-011823
- Data should default to saving in the following location: Documents > Agilent > [Run Date] > [File Name]