An issue was discovered in Smiths-Medical CADD-Solis...
Critical severity
Unreviewed
Published
May 17, 2022
to the GitHub Advisory Database
•
Updated Jan 27, 2023
Description
Published by the National Vulnerability Database
Feb 13, 2017
Published to the GitHub Advisory Database
May 17, 2022
Last updated
Jan 27, 2023
An issue was discovered in Smiths-Medical CADD-Solis Medication Safety Software, Version 1.0; 2.0; 3.0; and 3.1. CADD-Solis Medication Safety Software grants an authenticated user elevated privileges on the SQL database, which would allow an authenticated user to modify drug libraries, add and delete users, and change user permissions. According to Smiths-Medical, physical access to the pump is required to install drug library updates.