Electrolyte bucket and solution, pipettes, towels
- Once cap is roughly set, move over to the testing room
- Connect the cap by plugging connector end of the cap to the arm/holder
- Lock by turning the knob
You can use the DVI switcher to show testing room display to mirror the NetStation computer display
- Open NetStation > Session > Choose PsychoPy EEG
- In NetStation, go to Panel > Impedance > Measure Impedance
- This should display the map of the cap - If the entire map is gray, it possibly means the reference electrode is not properly in contact - Once reference point is set properly, you can see green and red labels - Bring in the electrolyte bucket
- Using pipettes, apply additional electrolytes between the sponge and scalp to fix the red labels
- Provide towels/tissues as necessary
- Switch the DVI display to PsychoPy computer
- Run an example PsychoPy code to try out for basic tasks (pyns_expv2.py)